|
| |
- Gkanatsios NA, Mardirossian G, Matsushita T, Karellas A, Rajeevan
N, Garcia M, Redus R, Fogarty M, Bacharach S, Brill AB. "A
portable Camac LabView nuclear medicine imaging system."Physica Medica
9(2-3): 229-232 (1993).

The use of intraoperative devices requires compact detectors and
portable imaging systems. In order to image the biodistribution of radiolabeled
monoclonal antibodies at surgery, Radiation Monitoring Devices (RMD) developed a small
(3" diameter) scintillation camera using a Hamamatsu position sensitive
photomultiplier tube (psPMT) coupled to a collimated NaI(Tl) scintillator. The
portable data acquisition system we developed and used to collect and display images was
coupled to a Macintosh IIfx computer via CAMAC-based nuclear electronics. The system
software has been written in LabView 2.2 (National Instruments Corp.). The design of
the system is presented along with an evaluation of its performance in planar and
tomographic image reconstruction. Software for data acquisition, planar and
tomographic image analysis and display was written entirely in LabView 2.2.

- Geiser WR, Huda W, Gkanatsios NA. "Effect
of patient support pads on image quality and dose in fluoroscopy." Medical
Physics 24(3): 377-382 (1997).

An investigation was performed of the changes in image quality and
patient dose as a result of increasing filtration for fluoroscopy performed under
automatic brightness control. Filtration was added either at the x-ray tube housing
(i.e., scatter-free geometry) or adjacent to a tissue equivalent phantom simulating the
patient (i.e., with-scatter geometry). Patient doses were expressed in terms of the
total energy imparted to patients simulated by either a 10 cm (i.e., pediatric) or 20 cm
(i.e., adult) acrylic phantoms. Changes in image quality were determined by
measuring the relative visibility of circular disks in a Leeds Test Object 10
contrast-detail phantom. In the scatter-free geometry, the addition of 4 mm Al
filtration reduced the energy imparted by 27% (10 cm phantom) and 20% (20 cm phantom).
In the with-scatter geometry, the corresponding reductions in energy imparted were
17% and 9% for the 10 and 20 cm phantoms, respectively. The visibility of low
contrast disks generally decreased as the thickness of the added aluminum increased but
the location of the added Al (i.e., with-scatter or scatter-free geometry) had no
significant effect on the resultant image quality. These results demonstrate that
the use of patient support pads with a thickness of about 4 mm Al will generally have an
adverse impact on fluoroscopic image quality and result in modest reductions (about 10%)
of adult patient doses.

- Gkanatsios NA, Huda W. "Computation of
energy imparted in diagnostic radiology" Medical Physics 24(4): 571-579
(1997).

Energy imparted is a measure of the total ionizing energy deposited
in the patient during a radiologic examination and may be used to quantify the patient
dose in diagnostic radiology. Values of the energy imparted per unit exposure-area
product, w(z), absorbed by a semi-infinite water phantom with a
thickness z, were computed for x-ray spectra with peak x-ray tube voltages
ranging from 50-140 kV and with added filtration, ranging from 1- 6 mm aluminum. For
a given phantom thickness and peak x-ray tube voltage, the energy imparted was found to be
directly proportional to the x-ray beam half-value layer (HVL) expressed in millimeters of
aluminum. Values of w(z) were generated for constant waveform
x-ray tube voltages and an anode angle of 12°, and were fitted to the expression
w(z)=a x HVL + b.
Fitted a and b parameters are provided
which permit the energy
imparted to be determined for any combination of tube voltage, half-value layer, and
phantom thickness from the product of the entrance skin exposure (free-in-air) and the
corresponding x-ray beam area. The results obtained using our method for calculating
energy imparted were compared with values of energy imparted determined using Monte Carlo
techniques and anthropomorphic phantoms for a range of diagnostic examinations. At
60, 80 and 120 kV, absolute values of energy imparted obtained using our method differed
by 8%, 10% and 12% respectively, from the corresponding results of Monte Carlo
computations obtained for an anthropomorphic phantom. The method described in this
paper permits a simple determination of energy imparted for any type of diagnostic x-ray
examination which may be used to compare the radiologic risks from differing types of
x-ray examinations, optimize imaging techniques with respect to the patient dose, or
estimate the patient effective dose equivalent.

- Gkanatsios NA, Huda W, Peters KR, Freeman JA.
"Evaluation of an on-line patient exposure meter in
neuroradiology." Radiology 203: 837-842 (1997).
PURPOSE: To assess the clinical performance and
usefulness of an on-line patient exposure meter installed on a neuroradiologic biplane
imaging system.
MATERIALS AND METHODS: A commercial on-line patient exposure meter was
installed on each plane of a biplane neuroradiologic imaging system. The meter
computed skin exposures on the basis of selected technique factors (tube potential and
current) and information about patient location relative to the x-ray tube.
Simulations were performed to measure the system accuracy with an angiographic
anthropomorphic head phantom with the skin exposures measured with an ionization chamber.
Skin doses were computed for 114 consecutive patients who underwent diagnostic and
interventional neuroradiologic procedures.
RESULTS: Agreement between measured and computed skin exposures in
fluoroscopy, plain radiography, and digital imaging was generally within 5% of the true
skin dose. For all fluoroscopic and radiographic procedures, total median skin doses
were 1.20 and 0.64 Gy for the frontal and lateral planes, respectively. in both
planes, patient skin doses resulted primarily from digital subtraction angiographic
acquisitions. In 29 (25%) patients, the skin dose exceeded 2.00 Gy, but no
radiation-induced deterministic effects were observed.
CONCLUSIONS: An on-line patient exposure meter can provide accurate
radiation skin dose data in patients undergoing diagnostic and therapeutic neuroradiologic
procedures

- Huda W, Gkanatsios NA. "Effective dose
and energy imparted in diagnostic radiology" Medical Physics 24: 1311-1316
(1997).

The patient effective dose, E, is an indicator of the
stochastic radiation risk associated with radiographic or fluoroscopic x-ray examinations.
Determining effective doses for radiologic examinations by measurement or
calculation is generally very difficult. By contrast, the energy imparted, e,
to the patient may be obtained from the x-ray exposure-area product incident on the
patient. As energy imparted is approximately proportional to the effective dose for
any given x-ray radiographic view, the availability of E/e ratios for
common radiographic projections provides a convenient way for estimating effective doses.
Ratios of E/e were obtained for 68 projections using E
and e values obtained from published dosimetry data computed using Monte Carlo
techniques on an adult anthropomorphic phantom. The average E/e
ratio for the 68 projections in adults was 17.8 ± 1.4 mSv/J whereas uniform whole body
irradiation corresponds to 14.1 mSv/J. The major determinant of E/e
ratios was the projection employed (body region irradiated and x-ray beam orientation) whereas the tube
potential and beam filtration were of secondary importance. Adult E/e
ratios may also be used to obtain effective doses to pediatric patients undergoing x-ray
examinations by application of a correction factor based on the patient mass.

- Huda W, Gkanatsios NA. "Radiation
dosimetry for extremity radiographs." Health Physics 75: 492-499 (1998).

The energy imparted, e,
to a patient undergoing an extremity
x-ray examination may be obtained from the dose-area product incident on the patient.
Values of energy imparted can be subsequently converted into the corresponding
effective dose, E, using an extremity specific E/e ratio.
In this study, an E/e ratio of 3.0 mSv/J was used to convert
values of energy imparted into the corresponding upper limit of adult effective doses for
all types of extremity examinations. A modification factor, based on the patient
mass, was employed to determine the corresponding extremity effective doses to pediatric
patients undergoing extremity examinations. Representative clinical technique
factors for six common extremity examinations (hand, forearm, elbow, ankle, tibia/fibula,
knee) were used to determine the dose-area product and the corresponding values of energy
imparted. For adult extremity x-ray examinations, values of energy imparted ranged
from 55 µJ to 920 µJ, with the energy imparted to 1-year-old patients being a factor
of about 20 lower. Upper limits of effective doses for adult extremity x-ray
examinations ranged from 0.17 to 2.7 µSv, whereas the corresponding doses to
1-year-old patients were about a factor of three lower.

- Rill LN, Huda W, Gkanatsios NA. "Viewbox
luminance measurements and their effect on reader performance." Academic Radiology
6: 521-525 (1999).
PURPOSE: To measure viewbox luminance values in a
teaching hospital, and to investigate the importance of viewbox luminance and viewing
conditions on low contrast detection performance.
MATERIALS AND METHODS: The luminance values of 632 viewbox panels in a
teaching hospital were obtained using a calibrated meter. Radiographic images of a
mammography contrast-detail phantom were viewed on viewboxes with luminance values ranging
from 860 nit to 3300 nit. Twelve radiologists reported the number of contrast detail
disks for each size (0.3 mm to 7 mm diameter) that were deemed to be visible for films
with optical densities ranging from 1.0 to 2.6. Radiologist performance in reading
low contrast phantom images was also studied as a function of room illuminance and image
masking.
RESULTS: The median luminance value was 1700 nit with 25 and 75 percentile
values of 1450 and 2150 nit, respectively. In general, low contrast visibility was
independent of viewbox luminance irrespective of film density or disk diameter. Low
contrast visibility deteriorated with removal of masking around the image as well as when
images were read in normal room lighting. The greatest deterioration in performance
was observed at the highest film densities and the smallest size disks.
CONCLUSIONS: Detection of low contrast features in radiographs is
relatively independent of viewbox luminance, but is significantly degraded by the presence
of stray light and by increased room illuminance.

- Grable RJ,
Ponder SL, Gkanatsios NA, Olivier P, Hall DJ, Zeng Y, Wake R. "Optical computed tomography for imaging the breast: first look."
Proceedings of SPIE Vol. 4082: 40-45 (2000).

The purpose of
the study is to compare computed tomography optical imaging with traditional
breast imaging techniques. Images produced by a computed tomography laser
mammography (CTLM™) scanner are compared with images obtained from mammography, and in some cases
ultrasound and/or magnetic resonance imaging (MRI). During the CTLM procedure,
a near infrared laser irradiates the breast and an array of photodiodes
detectors records light scattered through the breast tissue. The laser and
detectors rotate synchronously around the breast to acquire a series of slice
data along the coronal plane. The procedure is performed without any breast
compression or optical matching fluid. Cross-sectional slices of the breast
are produced using a reconstruction algorithm. Reconstruction based on the
diffusion theory is used to produce cross-sectional slices of the breast.
Multiple slice images are combined to produce a three-dimensional volumetric
array of the imaged breast. This array is used to derive axial and sagittal
images of the breast corresponding to cranio-caudal and medio-lateral images
used in mammography. Over 200 women and 3 men have been scanned in clinical
trials. The most obvious features seen in images produced by the optical
tomography scanner are vascularization and significant lesions. Breast
features caused by fibrocystic changes and cysts are less obvious. Breast
density does not appear to be a significant factor in the quality of the
image. We see correlation of the optical image structure with that seen with
traditional breast imaging techniques. Further testing is being conducted to
explore the sensitivity and specificity of optical tomography of the breast.

- Gkanatsios NA,
Huda W, Peters KR. "How does magnification affect
image quality and patient dose in digital subtraction angiography?"
Proceedings of SPIE Vol. 4320: 326-330 (2001).

Digital subtraction angiography (DSA) images were
obtained of a phantom containing 1 mm diameter vessels. The iodine
concentrations ranged from 5 to 50 mg/cc, which permitted the determination
of the detection threshold concentration of iodine. The source to
image receptor distance was 105 cm, and image magnification was varied
between 1.15 and 2.0. One experiment was performed at an input
exposure of 1 mGy per frame, and a second experiment was performed at 4
mGy per
frame. Surface (skin) doses were measured using an ionization chamber,
and the corresponding values of energy imparted were determined from the
exposure-area product. Increasing the radiation exposure by a factor
of four reduced the threshold iodine concentration by ~30%. The
average detection threshold iodine concentration for a magnification of 1.15
was 13.6 mg/cc. Detection performance improved linearly with
magnification, with an average value of 7.4 mg/cc at a magnification of
2.0. Values of energy imparted were essentially independent of
geometric magnification, whereas surface doses increased by a factor of four
when geometric magnification was increased from 1.15 to 2.0.
Magnification offers improved image performance at no additional patient
risk provided that surface doses do not exceed the dose threshold for
deterministic effects such as skin burns and epilation.

- Gkanatsios NA,
Huda W, Peters KR. "Adult patient doses in
interventional neuroradiology."
Medical Physics 29: 717-723 (2003).

We investigated the radiation doses to 149 adult patients who underwent interventional
neuroradiologic procedures, including 132 patients who had diagnostic imaging examinations and the remaining 17
patients who had therapeutic procedures. All procedures were performed on a biplane imaging system capable of
performing fluoroscopy and digital subtraction angiography (DSA) imaging.
The x-ray imaging system has a commercial patient dosimetry system which computes surface (skin) doses based on the
selected radiographic technique factors in both radiographic and fluoroscopic imaging modes. For each patient,
an assessment was made of the maximum surface dose received during the procedure, which is a predictor of the
possibility of inducing deterministic effects. Knowledge of the surface doses, beam quality and x-ray
cross sectional area permitted the total energy imparted to the patients to be determined. We computed
the corresponding effective dose to each patient, which is an estimate of their stochastic radiation risk,
using projection specific effective dose to energy imparted ratios. In general, doses for the lateral plane were
much lower than those observed for the frontal plane. For diagnostic imaging examinations, the median
surface dose for the frontal plane was 1.3 Gy, with a maximum surface dose of 5.1 Gy. For therapeutic procedures,
the median surface dose for the frontal plane was 2.8 Gy with a maximum surface dose of 5.0 Gy.
In interventional neuroradiology, surface doses could result in the induction of deterministic effects.
The median energy absorbed was 1.77 J during fluoroscopy, and 4.30 J for radiographic imaging, showing that most of
the patient exposure to radiation results from DSA imaging. For diagnostic imaging examinations, the median patient
effective dose was 22 mSv, with a maximum of 102 mSv. For therapeutic procedures, the median patient effective dose
was 51 mSv, with a maximum of 97 mSv. Patient effective doses in interventional neuroradiology are much higher than
those of other common types of diagnostic procedures, which employ ionizing radiation.

- Gkanatsios NA,
Huda W, Peters KR. "Effect
of radiographic techniques (kVp & mAs) on image quality and patient dose
in digital subtraction angiography."
Medical Physics 29: 1643-1650.

We investigated how varying the x-ray tube voltage and image receptor
input exposure affected image quality and radiation doses in interventional neuroradiologic imaging.
Digital subtraction angiography (DSA) images were obtained of a phantom with 1 mm diameter vessels containing
iodine at concentrations between 4.5 and 50 mg/cc. The detection threshold concentration of iodine was
determined by inspecting DSA images obtained at a range of x-ray tube voltages and input exposure levels.
Surface doses were generated using measured x-ray tube output data, and the corresponding values of energy imparted
were determined using the exposure-area product incident on the phantom. In one series of experiments,
the air-kerma at the image intensifier (X) was varied between 0.44 mGy per frame and 8.8
mGy per frame at a constant x-ray tube voltage of 70 kVp.
In a second series of experiments, the kVp was varied between 50 and 100 kVp, and the mAs adjusted to maintain a
constant exposure level at the image intensifier input surface. At a constant x-ray tube voltage,
the surface dose and energy imparted were directly proportional to the input exposure per frame used to acquire
the DSA images. Below 2.2 mGy per frame, the threshold iodine concentration
was found to be proportional to X-0.59, in reasonable agreement with the theoretical prediction for a quantum
noise limited imaging system. Above 2.2 mGy per frame, however, the threshold iodine
concentration was proportional to X-0.27, indicating that increasing the input exposure above this value
will only achieve modest improvements in image quality. At a constant image intensifier input exposure level,
increasing the x-ray tube voltage from 50 kVp to 100 kVp reduced the surface dose by a factor of 6.1, and energy
imparted by a factor of 3.5. The detection threshold iodine concentration was found to be proportional
to kVpn, where n was 2.1 at 1.1 mGy per frame, and 1.6 at
3.9 mGy per frame. Improvements in DSA imaging performance are best achieved by
reducing the x-ray tube voltage rather than increasing the exposure level, since this will generally minimize
patient doses.

- Gess D, Gkanatsios NA, Jones C. A
Revealing Look Into The Department Of Defense's Use Of Military Specifications.
Washington, DC: Manufacturing Studies Board, National Research Counsil (1990)
The Department of Defense (DOD) spends approximately $170 billion on
15,000,000 contracts each year. In these contracts, the DOD uses military specifications
(milspecs) to describe the products being purchased. Milspecs have recently come
under fire from Congress for being inefficient, costly, and wasteful. The DOD was
chastised by Congress for using milspecs that contained incorrect data, which led to
exorbitant expenditures on such items as a $640 toilet seat and $7400 coffeepot.
The Manufacturing Studies Board has commissioned this report to study the use of milspecs
and some of the problems associated with milspecs in order to help them develop a defense
manufacturing strategy for the Under Secretary of Defense (Acquisition), John A. Betti.
To begin the study, an investigation into the use of milspecs within the Defense
Acquisition Process was necessary. This background information was crucial in
evaluating the problems associated with the DOD's use of military and commercial
specifications.
Upon investigating how milspecs are used within the Defense Acquisition Process, it was
found that milspecs are blamed for being too long and complicated, and for containing
obsolete technology. In reference to their complexity, milspecs, by their nature,
are technical documents which describe the products the DOD procures. Contractors
and DOD engineers who actually use milspecs usually have no problem understanding the
technical language used. There are, however, instances where milspecs can be overly
complicated due to the inexperience of new writers.
In addition to being too complicated, milspecs in rapidly changing fields such as
semiconductors and computer software have been criticized for containing obsolete
technology. Even though milspecs are reviewed every five years, they can be updated
with Engineering Change Proposals.
Another problem occurs when writers do not examine all the specifications they reference
to ensure that they are updated and relevant to the product. Due to time constrains,
the writers often cannot examine all the references. Therefore, there may be some
outdated or unnecessary references included in the final specification.
Many people say the DOD can avoid the problems with milspecs, as in the case of products
such as the toilet seat, by buying directly from the commercial market (Defense Science
Board, 1986). The DOD's present policy is to buy commercial products as often as
possible. However, there are several limiting factors to this policy. Commercial
products are often less expensive and work well in a commercial environment, but become
unreliable in a military environment. The commercial products also are not
standardized with existing military equipment.
After analyzing these issues, this report states one major recommendation: the military
and commercial industrial bases should begin to integrate into one base. The
integration of the two industrial bases is a long term recommendation. To support
the integration in the near future, some short term recommendations that can be applied
presently are:
The Committee on Defense Manufacturing Strategy should be cautious
about accepting myths about milspecs as fact.
All new milspec writers should be required to take a
specification writing class which also explains how the documents are used in the Defense
Acquisition Process.
The Committee on Defense Manufacturing Strategy should investigate the
communication over specification changes within a contract between the DOD and
contractors.
Writers should be required to submit an Engineering Change
Proposal as soon as they see that a milspec is outdated.

- Gkanatsios NA. Bachelor Thesis (MQP):
A CAMAC LabView Macintosh Nuclear Medicine Imaging System.
Worcester Polytechnic Institute, Worcester, MA (1992).
This report, prepared at the Nuclear Medicine Department of the
University of Massachusetts Medical Center, focuses on the synthesis of a working nuclear
imaging system. A 3" diameter Hammamatsu position sensitive PMT/3 mm NaI(Tl)
scintillator is interfaced to a Macintosh IIfx computer via a CAMAC acquisition system.
An analog interface pulse height board selects wanted events and computes their
location. Square analog pulses representing the coordinates of each nuclear event
are sent to two ADCs, which pass digitized addresses to a special built module used to
strip the lower order bits from incoming information to 5, 6, 7, or 8 bits. The same
module also binds the two lateral addresses together into the low order 10, 12, 14, or 16
bits, respectively. Two inputs are also available for setting the 15th and 16th bits
high (e.g., for physiological triggers, or timing information). These data are then
passed to a data router, which controls the characteristics of the image. The data
router directs the information to two parallel histogramming memories, each of which can
buffer one 128x128, or four 64x64 histogramming images. A GPIB controller module in
CAMAC passes the data to a Macintosh IIfx computer (System 7.0). LabView 2 is used
as the programming environment for acquiring and archiving 128x128 static images, or
sequential four 64x64 frames without interframe dead time. List mode studies can
also be supported via two dual port memories. Software support for image analysis
and display in provided in LabView, along with C, FORTRAN, and Pascal support. The
system is designed for use as an intraoperative camera, and for small organ imaging.
The possibility of using the system in Tomography for three dimensional
reconstruction was also evaluated. Intrinsic resolution of the system is 300 micrometers,
and 2 mm resolution has been achieved with a high resolution parallel hole collimator.

- Gkanatsios NA. Master Thesis: Computation
of Energy Imparted in Diagnostic Radiology. University of Florida,
Gainesville, FL (1995).
Integral energy imparted to a patient is a better indicator of
radiological risk than the entrance skin exposure, since the former takes into account
both the area of exposure and x-ray beam penetration. However, data on energy
imparted are not readily available. The objective of this study was to generate
values of energy imparted per unit exposure per unit area of exposure as a function of
patient thickness and x-ray beam qualities (kVp, HVL). A semi-empirical model
developed by Tucker et al. was used to generate the x-ray spectra. X-ray
spectra were obtained for a range of kV, tube anode angles and x-ray beam filtration for a
tungsten-rhenium target. A sine function was used to model the x-ray tube kilovolt
waveform with a ripple factor ranging from 0% (i.e., constant potential) to 100% (i.e.,
single phase). X-ray spectra were filtered by varying the thickness of aluminum to
simulate x-ray beams encountered in clinical practice. For each x-ray spectrum the
free-in-air exposure was computed. Energy imparted to a semi-infinite, water
equivalent phantom from polychromatic x-rays was computed according to Boones
parameterization of energy imparted. Values of energy imparted per unit exposure per
unit area of exposure were computed for constant potential and a 12° anode angle as a
function of patient thickness and x-ray beam qualities. Differences of less than 3%
were found to be introduced by varying the kilovolt waveform ripple or the tube anode
angle. Values of energy imparted to a water phantom were compared to those obtained
for an anthropomorphic phantom as determined by Jones and Wall. Absolute values
agreed to within 10% for typical x-ray examinations. The agreement among relative
values as a function of kV and filtration was better than 5% between the two phantoms.
Specification of the x-ray beam kVp and half value layer are generally sufficient
to yield energy imparted per unit exposure data that have accuracy of better than a few
percent. Data are presented in tabular form which permits the estimation of the
energy imparted to patients, for the range of kVp and HVL normally encountered in
diagnostic radiology, by specifying patient thickness and x-ray beam qualities.

- Gkanatsios NA, Huda W. "Patient
effective doses in diagnostic radiology." Conference Proceedings p.:
28-30, 43rd Annual Scientific Meeting (Canadian Organization of Medical
Physicists), Charlottetown, PEI Canada (1997).
The effective dose is the best parameter for describing the amount of
radiation received by a patient undergoing any diagnostic x-ray examination.
Benefits of the effective dose include the ease of intercomparing doses associated with
diverse types of radiologic examination and the ability to compare patient doses with
natural background or regulatory dose limits. The major limitation of the effective
dose is the requirement of obtaining mean organ doses of the irradiated tissues in the
patient. A method is proposed which can be used to generate patient effective doses
using the selected radiographic technique factors (i.e., kVp/mAs), patient source to skin
distance and x-ray beam cross-sectional area. An algorithm based on computed x-ray
spectra is used to generate the beam output (mR/mAs) and half-value layer at the selected
tube voltage for a given tube anode angle and waveform ripple. The energy imparted e
to the patient is then obtained from the exposure area product, x-ray beam voltage,
half-value layer and patient thickness. Energy imparted is finally converted to an
effective dose E using published body/projection specific E/e
ratios. Examples are provided of how the effective doses may be generated for
representative radiographic examinations of the head, chest, abdomen and for the
extremities.

- Huda W, Gkanatsios NA, Botash RJ, Botash AS.
"Pediatric effective doses in diagnostic radiology." Conference
Proceedings p.:114-116, 44rd Annual Scientific Meeting (Canadian
Organization of Medical Physicists), London, Ontario Canada (1998).
Pediatric effective doses can be obtained for any radiologic
examination using the selected radiographic technique factors (kV/mAs), the exposure
geometry and the patient mass. The energy imparted e to the patient may be
computed from the exposure area product, x-ray tube voltage, half-value layer and patient
thickness. Values of energy imparted may be subsequently converted to an effective dose E
using published radiographic projection specific E/e ratios determined
using Monte Carlo techniques applied to anthropomorphic phantoms, with a correction
applied for the patient mass. Pediatric effective doses (head, chest, abdomen and
extremity) were computed for representative adult patients, as well as for pediatric
patients ranging from new born to 15 year old youths. Values of patient effective dose
were dependent on body size, selected technique factors as well as the type of
radiographic imaging equipment used, with no clear trends for effective dose with patient
age.

- Gkanatsios NA. Ph.D.
Dissertation: Patient Doses and Image
Quality In Interventional Neuroradiology. University of Florida,
Gainesville, FL (1995).

Diagnostic and therapeutic interventional neuroradiologic procedures involve imaging of catheter
manipulation and vascular anomalies of the brain and generally require extensive use of x-ray radiation.
Knowledge of the surface dose allows one to estimate the probability of inducing deterministic effects, whereas the
corresponding value of effective dose is related to the patient stochastic risk.
Modification of key imaging parameters (i.e., tube voltage, input exposure to the image receptor and geometric
magnification) impact on patient doses and image quality, with the latter being defined as the lowest concentration
of iodine in a vessel that may be visually detected in the radiographic image.
A dosimetry system was installed on a biplane neuroradiologic imaging system to determine the doses to patients
undergoing interventional neuroradiologic procedures.
The dosimetry system computed surface doses on the basis of selected technique factors and information about patient
location relative to the x-ray tube. The energy imparted to the patient, e,
was determined using the surface dose, x-ray beam quality (i.e., kVp and HVL), exposure area and thickness of the
patient and was converted into the corresponding value of effective dose, E. Values of surface dose and
E were obtained for 175 patients, consisting of 149 adults and 26 pediatrics.
Median values of surface doses to the head region were 1.2 Gy in the frontal plane and 0.62 Gy in the lateral
plane. Median values of the effective doses were 36 mSv for adult patients and 44 mSv for pediatric patients.
An acrylic phantom with 1-mm diameter vessels filled with iodine contrast was used to evaluate the effects of
varying imaging parameters on signal detection and patient doses during digital subtraction angiography.
Reducing the x-ray tube voltage offered the largest improvement in image quality for a given increase in patient
dose. Increasing the image intensifier input exposure beyond 250 mR/frame provided
very little improvement in image quality, and this II exposure level should not be exceeded in interventional
neuroradiologic imaging. A linear relationship was observed between magnification and threshold concentration,
which offers significant patient benefits when surface doses are not expected to exceed the threshold doses
for the induction of deterministic effects.

- Grable R, Gkanatsios NA, Ponder SL. "Optical
mammography" Applied Radiology 29: 18-20 (2000).
The field of biomedical optics, the use of light in
medicine, has seen rapid growth over the last decade. The development
of non-invasive medical optical imaging (MOI) modalities using harmless near
infrared (NIR) light is a major goal of biomedical optics research.
Arguably, the most interesting application of MOI is breast imaging, with
the potential to provide a safer, and possibly more effective alternative to
traditional x-ray mammography.

|